Gafchromic in vivo Dosimetry System
New Launch
Our new gafchromic™ film-based in vivo dosimetry system brings innovation to the forefront of radiation monitoring, providing unparalleled precision in real-time dose verification
A Versatile Solution
Transform your radiation therapy monitoring process with:
-
Straightforward setup and automated calibration process.
-
Automated roi capture, documentation and reporting features.
-
Safeguarding alerts built into the software module for user experience & towards maximizing precision
-
Single use (gafchromic™ film based): pnt-dos™ device can be used for both high and low dose photon and electron measurements making it a long-term, cost-efficient dosimetry tool.
Key Features and Benefits
-
Near real-time in vivo dose solution
-
Precisely measures ionizing radiation surface doses immediately after delivery, help ensuring compliance with prescribed treatment plans.
-
Provides immediate feedback for accurate dose delivery during radiation therapy.
-
-
Gafchromic™ film technology
-
Industry-leading radiation-sensitive films for exceptional sensitivity and accuracy.
-
No need for external power or cables—completely passive operation.
-
Eliminates electronic interference in sensitive treatment environments.
-
-
Non-invasive and patient-friendly
-
Compact Gafchromic film-based technology flexible, biocompatible design that conforms easily to the patient’s anatomy.
-
Seamless placement ensures patient comfort during prolonged procedures
-
pnt-dos™ Device Type (Gafchromic Film based)
​
-
Measurement range: up to 400 cgy
-
Accuracy: +/- ±10 cgy for low dose [doses <80 cgy] & ±5% for high dose [80cgy to 400 cgy].
-
Size: 1cm × 1cm active area (region of interest)
-
Engineered to deliver precise measurements that empower clinicians to confidently fine-tune radiation doses.
-
energy dependence: energy dependence analysis revealed deviations of up to ±5% (results during validation) for megavoltage photon and electron energies.
-
Angular dependence: maximum variation ±3% (results during validation) when the beam passed through the posterior oblique side of the device.
-
Daily qc/reproducibility tests confirmed optical scanner
system constancy. -
Time-resolved calibration enabling near real-time readouts
How it works






a)scanner qa interface b)the new in-vivo device size compared to a penny c)calibration interface showing a preview of the auto-detected ROls, user input of irradiation time and time intervals for auto-scanning d) an example of the calibration set with low/high calibration curves at each time point with a slider bar to view all calibration curves in the set e) in-vivo measurement reading interface showing user selection of calibration curve and user input of irradiation time
​​
Regulatory Compliances
-
ISO 13485 certified manufacturing
-
FDA registered manufacturing location
-
CE marked
Additional Information
Product part and description
987963 Gafchromic in-vivo dosimetry system filmQApro v8 with scanner
990186 Gafchromic in-vivo dosimetry system FilmQA pro 8 w/o scanner
987961 Gafchromic pnt-dos device (53 devices per box)